¹³¹I sodium iodide · MIRD schemaPre-Treatment Basic Dose Calculation (77300)

I-131  T½,phys 8.02 d (192.6 h)
β⁻ mean 182 keV · γ 364 keV (81%)

Estimate only. This scales the ICRP Publication 128 reference-adult dose coefficients for ¹³¹I-iodide, to one patient's size and renal function with independent physician verified model calculations for detection of error before treatment (independent verification). In thyroid cancer the blocked-thyroid coefficients are used (no uptake assumed, only patient factors) and the blood dose uses the EANM formula with a residence time modelled from renal function. In hyperthyroidism the coefficients are interpolated to the measured uptake and the thyroid dose uses the Marinelli formula with the gland mass, uptake and effective half-life. It is not a substitute for measured pre-therapy dosimetry. Organ limits use beta (radiopharmaceutical-therapy) values where available, then external-beam values.

Patient & regimen

kg
cm

* Required. The activity fills in from the treatment intent; you can overwrite it. Thyroid cancer doses use only patient factors (weight, height and GFR) to adjust the classic ICRP 128 blocked-thyroid coefficients, so they can be calculated before any uptake measurement. No remnant uptake is assumed; iodine-avid remnant or metastases would retain activity longer and raise the whole-body, blood and marrow doses. With hormone withdrawal, a baseline GFR (measured on levothyroxine) is reduced by 19% for the hypothyroid state (Coura-Filho et al. 2015: ⁵¹Cr-EDTA GFR fell from 94 to 76 mL/min/1.73 m² during withdrawal and did not change with rhTSH); choose "During hypothyroidism" if the GFR was measured after withdrawal, so it is used as entered.

BSA
–
Abs. GFR
–
Clearance ratio
–
Reference patient & limits
kg
cm
mL/min
g
h
L/kg

Dose limits

OrganLimit (Gy)Basis

Limits follow this order of preference: beta (radiopharmaceutical-therapy) limits, then external-beam limits. Beta (RPT) values from Wahl et al. 2021 (Normal-tissue tolerance to radiopharmaceutical therapies, the knowns and the unknowns, J Nucl Med 2021;62(Suppl 3):23S–35S) are used for red marrow, kidneys and lungs, with the Benua 2 Gy blood limit for thyroid cancer; external-beam values are used for the salivary glands, liver and urinary bladder wall (Wahl 2021; QUANTEC for the submandibular glands) and the stomach wall (Emami 1991). In hyperthyroidism the thyroid has a target dose, not a limit. Treat a flag as a prompt for review, not as a contraindication. Clear a value to remove that organ's limit.

Administered activity
–
–
Blood (EANM) · this administration
–
–
Whole-body retention at 48 h
–
–

Absorbed dose by organ

OrganGy/GBqMass ×TIA ×Absorbed dose (Gy)vs limit / target
CalcICRP ref

Doses are for one administration of the activity entered. Calc is this patient's adjusted dose; ICRP ref is the ICRP 128 reference-adult coefficient × the same activity, with no patient adjustment. In thyroid cancer the classic blocked-thyroid coefficients are used (no uptake assumed), adjusted only for organ mass and renal function, so no uptake or kinetic measurement is needed. In hyperthyroidism, ICRP 128 gives coefficients for a blocked thyroid and for low, medium and high thyroid uptake, taken here as 0, 15, 35 and 55% (the ICRP 53 uptake levels) and interpolated linearly to the patient's uptake. Each coefficient is split into an iodide part (the blocked-thyroid value × (1 − uptake)), which is cleared by the kidneys and scaled for GFR, and a thyroid-derived part (the rest), which is scaled by the patient's thyroid or remnant Teff relative to the reference. The parotid and submandibular glands use the ICRP 128 salivary gland value, and both kidneys and lungs use the ICRP whole-organ values.

MIRD calculation by organ

Each organ's dose worked through step by step with this patient's numbers. Ã/(f·A₀) is the time-integrated activity per unit of activity taken up by the organ (mono-exponential: 1.443 × Teff).

Method & reference coefficients

MIRD absorbed dose is the time-integrated activity in each source organ times the S value. Iodide that is not taken up by the thyroid is excreted by the kidneys (biological half-time about 6 h in the ICRP model), so renal function sets how long it stays in the body. Thyroid and remnant doses use the Marinelli formula, which is the MIRD self-dose for a gland that takes up a fraction of the activity and clears it with an effective half-life.

D(organ) = Ã × S ≈ A₀ × τ × S

D_pt/A₀ = M × [D_iodide × TIA + D_thyroid-derived × T_eff,thy,pt / T_eff,thy,ref]

D_iodide   = D_blocked × (1 − U)        U = uptake fraction
λ_bio,pt   = λ_bio,ref × [(1 − w) + w × r]     r = GFR_pt / GFR_ref
TIA        = T_eff,pt / T_eff,ref

Thyroid (Marinelli):  D (Gy) = A₀ (MBq) × U (%) × T_eff (d) / (24.67 × m (g))
Blood (EANM SOP):     D/A₀ (Gy/GBq) = 108 × τ_blood (h/mL) + 0.0188 / wt^(2/3) × τ_WB (h)

w is the fraction of each organ's biological clearance assumed to depend on renal excretion. Lower GFR keeps iodide in the blood longer, lengthening Teff and raising the dose. The ratio r is clamped at the minimum set above.

OrganBlockedLow (15%)Medium (35%)High (55%)Teff,ref (h)Renal wMass scales by

Reference coefficients (Gy/GBq, numerically equal to mGy/MBq) are the ICRP Publication 128 adult values for ¹³¹I-iodide (oral), as reproduced in the sodium iodide (¹³¹I) therapy capsule product information. Teff,ref is the effective half-life of the iodide pool (ICRP blood biological half-time 0.25 d with physical decay); renal weights are typical values. Check them against your institutional data.

Complete report

Inputs, dose summary, organ table, step-by-step MIRD calculation for every organ, method and references.